pH Meter
Our pH meter is the Thermo Scientific Orion Lab Star PH111 pH/mV bench meter.
ThermoFisher provides a good video for calibration:
Calibration
Don’t break the electrode
Be sure electrode is ALWAYS in liquid. Dont let it dry out!
Calibrate frequently for best results.
Materials
Protocol
- Remove storage bottle from electrode.
- Rinse electrode with distilled water to remove any salt precipitate. Blot gently with kimwipe to remove excess water.
- Ensure electrode is filled enough. If not, refill with filling solution (3M KCl) into fill hole.
- Press the “Cal” key to start the calibration.
- Rinse the pH electrode with with buffer and then place into the ph buffer.
- Buffers can be done in any order.
- Ensure buffers are the same temperature as the sample to be measured.
- Wait for the pH value to stabilize.
- When the reading is stable, the checkmark icon is shown, and the reading is solid.
- Once the reading is stable, accept or edit the displayed pH buffer value. Press the “Enter” key to accept.
- Repeat steps 5-7 for the next calibration buffers.
- To save and end the calibration, press the “Cal” key.
- The meter will display the calculated slope value and then proceed to the main measure mode.
Measure pH
- Calibrate before measuring.
- Rinse electrode with distilled water. Blot gently with kimwipe to remove excess water.
- Place the electrode into the sample, submerging the measurement sensor and reference junction.
- Allow time for the reading to stabilize.
- The meter displays the stopwatch icon while the reading is stabilizing.
- The meter displays the checkmark icon when the reading is stable.
- Note the displayed measurements.
- Remove electrode from sample.
- Repeat steps 2-6 for all samples.
Storage
Short-Term Storage (up to one week)- Soak the electrode in the pH electrode storage solution. Cover the fill hole.
Long term storage (1+ week)- Rinse off any salt buildups with distilled water and remove any membrane/junction deposits. Fill the chamber with filling solution and securely cover the fill hole. Fill storage bottle with pH electrode storage solution and insert the electrode into the storage bottle.